X-Ray Generator High Voltage Power Supply50kV to 60kV

If you would like to learn more about this product: Please Contact Us

X-Ray Generator Power Supply

The XRG is a superior performance high voltage X-Ray generator power supply that provides a dependable source for tomography systems. Genvolt’s high frequency X-ray generator is a portable high voltage module. Subsequently allowing it to be integrated with many variants of X-Ray tube manufacturer’s products.

The X-Ray Generator Manufacturing Standards

Genvolt are a world leader in the design, manufacturing, and supply of high voltage power supplies. We combine our knowledge with class leading quality and robust builds. As well as that we appreciate that a consistent high voltage generator is an important element of an X-ray generation system. Thus, the XRG has been designed and manufactured to our exacting and impeccable standards. Consequently, each component has been engineered and integrated to withstand the rigours of use and high frequency interference.

Main X-Ray Generator Features

The XRG is small enough to fit on an A5 piece of paper. As a result, it weighs only 2Kg, making it highly portable. Undoubtedly these factors allow for easy incorporation into current or new equipment projects. Coupled with its compactness the unit is highly capable of meeting and exceeding expectations. Particularly with its output voltage range of 50kV to 60kV and filament current of 0-3.5A. In addition to these features, we can supply the XRG with either positive or negative polarity remote programming is available a DB15 interface.

If you need higher output voltages for X-Ray, then look at our Pegasus P1800.

X-Ray Generator Applications

The Genvolt XRG portable X-Ray generator is an ideal primary power source for an array of X-Ray imaging devices, be it analytical, industrial, or medical.

In brief the XRG has application uses in:

  • Analytical X-Ray
  • Automated Test Equipment (ATE)
  • Baggage Inspection
  • Biological Irradiation
  • Body Scanner
  • Bone Densitometry
  • Electron Spectroscopy
  • Electrophoresis
  • Electrostatics-Electrospinning
  • Elemental Analysers
  • Energy Dispersive X-Ray Fluorescence (EDXRF)
  • Equine
  • Explosive Detection (EDS)
  • Fill Level Inspection
  • Food Beverage Inspection
  • Industrial X-Ray
  • Irradiation
  • Industrial CT (Computerised tomography)
  • Liquid Inspection
  • Mammography
  • Mass Spectrometry
  • Non-Destructive Testing (NDT)
  • PCB Inspection
  • Production Processes / Control
  • Sterilization / Irradiation
  • Surgical-C-Arm
  • Thickness Gauging
  • Veterinary
  • X-Ray Diffraction (XRD)
  • X-Ray Fluorescence (XRF)
Technical Specifications for Genvolt XRG X-Ray Generator

There are multiple specifications and models, from 50kV to 60kV.
Positive & negative polarity can be selected (please specify when ordering).

Output Current 0-1mA
Input +24Vdc ±10%. Maximum Current 5A
Output Properties
High Voltage 50kV to 60kV
Beam Current 0-1mA
Filament Voltage Limited to 5V
Filament Current 0-3.5A
Programming Local and remote analogue communication interface
Protection Arc and short circuit
Ripple and Noise Better than 0.1%
Power Regulation Rate Better than 0.1%
Load Regulation Rate Better than 0.1%
Stability 30 min after starting up, less than every 8 hours 0.05%
Temperature Coefficient ≤25ppm/°C
 Local Output Control

Voltage & current are continuously adjustable in the full range, via an analogue remote control.

Environment
Operating Temperature -10 to 40°C. Storage -40 to 85°C
Humidity 10% to 90% relative humidity, non-condensing
Cooling Forced air cooling. The air inlet is through the front panel and the outlet is on the rear panel.
 Dimensions
HV Module (L x W x H) Metric 205mm x 104mm x 75mm
HV Module (L x W x H) Imperial 8.07" x 4.09" x 2.95"
Weight Approximately 2Kg (4.4lbs)
 Mechanical Dimensions

 X-Ray Generator Model XRG01

For local control, connect pin 7 to pin 8 (voltage) and pin 10 to pin 11 (current). Adjust the voltage and current setpoints via the potentiometer openings in the top cover that are labelled xxx and yyy. Power supply in the upper panel has a circular opening.


Adjust the internal potentiometer for setting voltage and current.

DB15 Pin Details

Pin Number Type Value
1 Ground Ground
2 Voltage Display 0-10Vdc full scal, Zout = 1kΩ
3 Current Display 0-10Vdc full scal, Zout = 1kΩ
4 External Interlock Short to PIN15, High Voltage Open
5 +10Vdc Reference Maximum Current 1mA, Voltage +10Vdc
6 Filament Current Display 1V=1A, Potentiometer Adjustment
7 Voltage Remote Control Input 0-10Vdc Full Scal, Zin = 10MΩ
8 Voltage Local Control Output 0-10Vdc, Potentiometer Adjustment
9 Filament Limit Setting Display 1V=1A, Potentiometer Adjustment, Zout = 10kΩ
10 Current Remote Control Input 0-10Vdc Full Scal, Zin = 10MΩ
11 Current Local Control Output 0-10Vdc, Potentiometer Adjustment
12 Idle Optional Safety Interlock Wiring
13 Idle Optional Safety Interlock Wiring
14 Filament Preheat Setting Display 1V=1A, Potentiometer Adjustment
15  +24V Ground +25V Ground

Remote Control Interface

Standard USB, RS232, Ethernet digital interface can control the power supply through GUI program. This makes it easy to operate and saves time. When the voltage and current settings are 0 at the same time, this is local control and can be used for display. When the voltage and current are not equal to 0, it is controlled by the software, and the power supply can be set and displayed.

1 TX+ 5 NC
2 TX- 6 RX-
3 RX+ 7 NC
4 NC 8 NC
Genvolt X-Ray Generator and High Voltage Power Supplies

In conclusion, Genvolt are a world leading manufacturer and distributor of efficient, high quality, high voltage power supplies. Likewise, our class leading portfolio of x-ray generator power supplies can support a wide variety of demanding operations. Designed exclusively by Genvolt, using the latest in electronics and high voltage techniques the XRG series are cost-efficient and sturdy. Last but not least they are all built to our exacting standards.